List of required parts
1
Speech processor with Angled Battery Pack
2
MLxi or MicroMLxS
3
MicroLink CI S
4
Red cable
Fitting with the user
1. Turn all equipment Off.
2. Connect the red cable to the MicroLink CI S and to the
back of speech processor. Make sure the cable is attached
correctly to the processor. Please check the processor user
guide.
3. Plug the FM receiver into the MicroLink CI S. When using
MicroMLxS, switch it to the double green dot position.
4. Turn the volume of the MicroLink CI S to the minimum.
5. Turn On the MicroLink CI S, speech processor and FM
transmitter.
6. Gradually increase the gain of the MicroLink CI S to a
comfortable level.
7. Determine speech recognition ability in quiet with the
FM signal. Ensure that results are similar to implant
alone condition. MicroLink CI S gain may need further
adjustments.
8. Evaluate FM benefit as well as the appropriateness of FM
settings, through speech-recognition assessment in the
clinic and standardized questionnaires completed by the
recipient, family members and teachers.
Listening check
The best way to listen to the FM signal is to plug the
MicroLink CI S with the blue patch cable into a loudspeaker.
Make sure there is a good battery inside the MicroLink CI S.
Then turn On the MicroLink CI S and FM transmitter.
By speaking into the transmitter’s microphone you should
hear your voice through the speaker.
2 3
41
MED-EL DUET1 / DUET2 / OPUS1 / TEMPO+
with angled battery pack
Picture shows OPUS1