Advanced Bionics PSP / Clarion S-Series /
Clarion 1.2
List of required parts
Pre-Fitting
Set audio mixing ratio to 50/50
Note: the audio mixing feature is not available for Clarion
S-Series and Clarion 1.2 processors.
Fitting with the user
1. Turn all equipment Off.
2. Connect the specific patch cable into the speech processor
and the MicroLink CI S.
3. Plug the FM receiver into the MicroLink CI S. If MicroMLxS
is used, move the switch to the double green dot position.
4. Turn the volume of the MicroLink CI S to the minimum.
5. To avoid transient sounds, turn On the FM transmitter,
then MicroLink CI S, then speech processor in that order.
6. Gradually increase the gain of the MicroLink CI S to a
comfortable level.
7. Determine speech recognition ability in quiet with the FM
signal. Ensure that results are similar to implant alone
condition. MicroLink CI S gain may need further adjust-
ments.
8. Evaluate FM benefit as well as the appropriateness of FM
settings through speech-recognition assessment in the
clinic and standardized questionnaires completed by the
recipient, family members and teachers.
Note
For the Clarion S-Series and Clarion 1.2 processors, the
processor microphone will be MUTED when an external device
such as the MicroLink CI S is plugged in
Listening check
The best way to listen to the FM signal is to plug the
MicroLink CI S with the blue patch cable into a loud speaker.
Make sure there is a good battery inside the MicroLink CI
S. Then turn On the MicroLink CI S and FM transmitter.
By speaking into the transmitter’s microphone you should
hear your voice through the speaker.
Platinum Sound
Processor (PSP) Clarion S-Series Clarion 1.2
1
MLxi or MicroMLxS MLxi or MicroMLxS MLxi or MicroMLxS
2
Blue-red short cable Blue short cable Black short cable
3
MicroLink CI S MicroLink CI S MicroLink CI S
4
PSP Clarion S-Series SP Clarion 1.2 SP
4
2
1 3
Picture shows PSP